Delta Elektronika SM6000 SERIES Laboratory Power Supply
The Delta Elektronika SM6000 Series is a high-power, precision DC power supply, engineered to meet the needs of demanding industrial, laboratory, and research applications. Delivering up to 6000W of output, with voltage options ranging from 15V to 600V and current capacities up to 400A, the SM6000 Series ensures reliable performance for diverse ...
Read MoreThe Delta Elektronika SM6000 Series is a high-power, precision DC power supply, engineered to meet the needs of demanding industrial, laboratory, and research applications. Delivering up to 6000W of output, with voltage options ranging from 15V to 600V and current capacities up to 400A, the SM6000 Series ensures reliable performance for diverse and critical operations. Its robust 4U design supports 19” rack mounting or standalone setups for maximum versatility.
Key features include Master/Slave parallel and series operation for high-power configurations, ultra-low ripple and noise for interference-free performance, and comprehensive protection against overloads and short circuits. The SM6000 Series offers flexible programming with optional high-speed functionality, digital encoders for precise adjustments, and two-quadrant operation for source and sink capabilities. Temperature-controlled fans ensure quiet and efficient cooling for continuous operation under load.
Ideal for applications such as automotive battery simulations, solar inverter testing, aerospace systems, and plasma chamber operations, the SM6000 Series is designed to handle the most challenging tasks with precision. Fully compliant with safety and EMC standards (cTUVus, EN60950, EN61010), it guarantees reliable and safe operation across various environments.
Choose the Delta Elektronika SM6000 Series Laboratory Power Supply for a powerful, adaptable, and precision-engineered solution to support your advanced technological needs with confidence.
Show LessFeatures
- 6000W Laboratory Supply
- Ranges 0-15V to 0-600V and 0-10A to 0-400A
- Designed for long life at full power
- Excellent dynamic response to load changes
- Protected against all overload and short circuit conditions
- EMC surpasses CE requirements for low emissions and high immunity
- Low audible noise - fans are temperature controlled
- Master/Slave parallel and series operation with voltage and current sharing
- High power system configuration from multiple units
- 19" rack mounting for for laboratory use (feet included)
- Stacking is allowed & space between units is not required
Specifications
| INPUT | |
|---|---|
| Input voltages | 342 - 528Vac 3Ph |
| Input Frequency | 48 - 62Hz |
| Power Factor | 400V 3Ph, full load 0.98, with 100% load 0.97 with 50% load |
| Input Current | 10.2A rms (400Vac 3ph) |
| Internal Fuse | 15 AT |
| Standby input power | 55W Vo = 0 110W Vo = max |
| Turn on delay | 200ms (after mains switch on |
| Inrush Current | 20A (electronic limit) |
| Phase Loss | Output shuts down in case of phase loss |
| Hold-up time | 6ms 100% load Vin=3x380Vac 15ms 50% load Vin=3x380Vac |
| OPERATING | |
|---|---|
| Series Operation | Max total voltage 600V Master / Slave operation yes |
| Parallel Operation | Max total current - No limit Master / Slave operation - max 3 units (including master) |
| Remote Sensing | Max volt, drop per load lead: 2V |
| Potentiometers & Encoders | Standard: Front panel control with knobs Resolution: 0.03% |
| Screwdriver adjustment | Option P001 |
| Digital encoders | Option P002 |
| Remote Shutdown | with +5V, 1mA or relay contact |
| Interlock | Contact at rear panel |
| Indicators | Voltage meter, Amp meter, AC-Fail, DC-Fail, Over Temperature, Power Sink Overload, Remote Shutdown, Remote-CV, Remote-CC, Output On,CV-limit, CV- and CC- mode |
| Controls | Mains on/off switch, CV and VV potmeter, CV and CC limit potmeter, Display Settings button, Display limits button, Remote/Local button, Output On/off button, Front panel lock button |
| Efficiency | 87% Typical |
| Isolation | Input - Output: 3750Vrms(1 min) Creepage clearance; 8mm Input - Case: 2500Vrms Output - Case: 600VDC |
| MTBF | 500,000hrs |
| OUTPUT | |
|---|---|
| Output Voltage Range | 0-15V to 0-600V Auto ranging models See Selection Table |
| Output Current Range | 0-10A to 0-400A Auto ranging models See Selection Table |
| Output Power Range | 0 to 6000W See Selection Table |
| STANDARDS AND APPROVALS | |
|---|---|
| Safety | cTUVus / EN60950 / EN61010 |
| Power Supply Standard | EN61204-3 Emission residential light industrial environment CISP22 Class B Immunity - industrial environment |
| General Emission | EN61000-6-3 residential, light industrial environment EN55022B EN61000-6-2, industrial environment |
| ANALOG PROGRAMMING | |
|---|---|
| Programming Inputs: CV | Input Range: 0 - 5V Accuracy: +/-0.2% Offset: -0.1...+1.3mV (on 5V) Temp. coeff. offset:: 10uV/C Input Impedance: > 1Mohm |
| Programming Inputs: CC | Input Range: 0 - 5V Accuracy: +/-0.5% Offset: 0...+2.2mV (on 5V) Temp. coeff. offset:: 50uV/C Input Impedance: > 1Mohm |
| Monitoring Outputs: CV | Input Range: 0 - 5V Accuracy: +/-0.2% Offset: -1.1..0mV (on 5V) Temp. coeff. offset:: 3uV/C Output Impedance: > 2 ohm /max 4mA |
| Monitoring Outputs: CC | Input Range: 0 - 5V Accuracy: +/-0.5% Offset: -1.1..0mV (on 5V) Temp. coeff. offset:: 60uV/C Output Impedance: > 2 ohm /max 4mA |
| Reference voltage | Vref: 5.114V +/-15mV (Ro = 2ohm, max 4mA) TC: 20ppm |
| +12V output | Vo: 12V+/-0.23V Imax: 0.2A Ro: 3ohm |
| ENVIRONMENTAL | |
|---|---|
| Operating Temperature | -20°C to +50°C (100% load) Derate output to 75% at 60°C |
| Temperature coefficient | 0.02%./°C |
| Storage temperature | -40°C to 85°C. |
| Thermal Protection | Output shuts down in case of insufficient cooling |
| Humidity | max 95% RH, non condensing up to 40°C max 75% RH, non condensing up to 50°C |
| MECHANICAL | |
|---|---|
| Mounting | Stacking of units allowed, air flow is from left to right. |
| Input Connector | Screw terminals for cable 2.5 - 4.0mm2 3 phase + earth (no neutral required) |
| Output Terminals | 0-15V: M12 bolts 0-30V, 0-45V, 0-60V,0-70V M10 bolts 0-120V, 0-300V, 0-600V; M8 bolts |
| Programming Connector | 15 pole D-connector at rear panel (FEMALE) |
| Enclosure | IP20 Protection |
| Dimensions | 177x 443 x 500mm Behind front panel (feet removed) 177 x 483mm, Front panel (19" 4U) |
| Weight | 27kg |
Selection Table
| OUTPUT | SM 15-400 | SM 30-200 | SM 45-140 | SM 60-100 | SM-70-90 | SM-120-50 | SM 300-20 | SM 600-10 |
| Output Voltage Range | 0 - 15V | 0 - 30V | 0 - 45V | 0 - 60V | 0 - 70V | 0 - 120V | 0 - 300V | 0 - 600V |
| Output Current Range | 0 - 400A | 0 - 200A | 0 - 140A | 0 - 100A | 0 - 90A | 0 - 50A | 0-20A | 0-10A |
| Meters | 3.5 digit | 3.5 digit | 3.5 digit | 3.5 digit | 3.5 digit | 3.5 digit | 3.5 digit | 3.5 digit |
| Scale voltage | 0 - 15.00 V | 0 - 30.0 V | 0 - 45.0 V | 0 - 60.0 V | 0 - 70.0 V | 0 - 120 V | 0 - 300 V | 0 - 600 V |
| Scale current | 0 - 400 A | 0 - 200.0 A | 0 - 140.0 A | 0 - 100.0 A | 0 - 90.0 A | 0 - 50.00 A | 0 - 20.0A | 0 - 10.0 A |
| Accuracy | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit | 0.5% + 2 digit |
| PROGRAMMING SPEED STANDARD VERSION(Resisitive Load) | SM 15-400 | SM 30-200 | SM 45-140 | SM 60-100 | SM 70-90 | SM 120-50 | SM 300-20 | SM 600-10 | |
|---|---|---|---|---|---|---|---|---|---|
| Rise time (10 - 90%) Output voltage step | 0 - 15V | 0 - 30V | 0 - 45V | 0 - 70V | 0 - 120V | 0 - 300V | |||
| Time, (100 % load) | 7 ms | ||||||||
| Time, (10 % load) | 7 ms | ||||||||
| Fall time (90 - 10%) Output voltage step | 15 - 0 V | 30 - 0 V | 45 - 0 V | 70 - 0 V | 120 - 0 V | 300 - 0 V | |||
| Time, (100 % load) | 7 ms | 7 ms | 87 ms | 8 ms | 7 ms | 11 ms | |||
| Time, (10 % load) | 32 ms | 58 ms | 29 ms | 82 ms | 39 ms | 91 ms | |||
| Programming bandwidth Small signal | 50 Hz | ||||||||
| Large signal,(100 % load) | 50 Hz | ||||||||
| Large signal,(10 % load) | 5 Hz | ||||||||